Siemens SOMATOM Emotion syngo CT 2007E Operator's Manual page 118

Table of Contents

Advertisement

Software Related Safety Information
A.4-24
syngo CT 2007E
02/2008
3D
Measurements
On the 3D task card, you have tools for region evaluation, meas-
uring distances, lengths, and angles, and for pixel evaluation.
CAUTION
Measurements in projected images!
Wrong diagnosis possible.
A . 4
Do not use measurements in projected images for diagnos-
tic purposes.
CAUTION
Use of pixel lens in the Viewer and in the 3D card.
The pixel lens may display different values.
A . 4
Be aware of differences in pixel lens measurements.
The pixel lens measurement in Viewer considers a rectan-
gle around the clicked point, whereas measurement in the
3D card considers a cube surrounding the clicked point.
Safety
Emotion Operator Manual
A . 4
A . 4
A . 4
A . 4
A . 4
A . 4
A . 4
A . 4
A . 4
A . 4
A . 4
0 . 0
0 . 0

Advertisement

Table of Contents
loading