Quality Assurance
Performing the homogenei-
ty test
syngo CT 2007E
02/2008
With this test, you measure the homogeneity of the CT values in
5 regions of the water phantom.
F . 3
Press the Start key on the control box.
Scanning is started.
F . 3
After the previous test has been completed, the Homogeneity
i
F
. 3
test is started automatically.
An image of the water phantom is recorded. 5 ROIs are marked
in the image, one central and four peripheral.
In addition to the ROIs, the mean values and standard devia-
tions of the CT values are displayed.
The difference between the mean values of the central ROI and
the peripheral ROIs is displayed.