Safety
syngo CT 2007E
02/2008
CAUTION
User enters wrong values for the calibration factors!
Wrong Calcium Scoring results may lead to wrong diagnosis.
A . 4
Only use correct calibration factor values, check the input
carefully.
CAUTION
The scanner may not be correctly identified by Calcium Scoring
during image load when using "remembered" calibration fac-
tors!
A . 4
Wrong calibration factors may lead to wrong results.
Carefully check the suggested factors and make sure that
they belong to the used scanner, before accepting the val-
ues.
Software Related Safety Information
Emotion Operator Manual
A . 4
A . 4
A . 4
A . 4
A . 4
A . 4
A . 4
A . 4
0 . 0
A.4-31
0 . 0
0 . 0