SPAN 1 calibration
5.
set up in TigerPC) are displayed along with a 30 second countdown. Attach the SPAN 1 gas
using the supplied Calibration Adaptor and press Enter to start the SPAN 1 countdown.
At the end of the countdown a tick '✓' will appear, indicating that the SPAN 1 has been
6.
accepted. Press Enter. For a two-point calibration, the calibration procedure is complete.
SPAN 2 calibration
7.
SPAN 2 are displayed (previously set up in TigerPC) along with a 30 second countdown.
Attach the SPAN 2 gas and press Enter to start the SPAN 2 countdown.
To skip SPAN 2, press Esc and then 'skip' to complete a SPAN 1 only calibration and return
to the main running screen.
At the end of the count down a tick '✓' will appear, indicating that the SPAN 2 has been
8.
accepted. Press Enter again to complete the calibration procedure.