then a 0.3L/min is recommended, a flow adaptor should be used to avoid over-pressurising the
instrument (sold separately). Please ensure you are familiar with the entire calibration procedure
before attempting to calibrate your TIGER Select.
The calibration of your TIGER Select must be carried out in a clean air
environment. Ensure all parts of the Calibration Kit are available and ready for
use.
Never calibrate the Zero with the Span gas connected.
PID Custom Calibration Procedure
From the main running screen, select the 'CAL' icon on key B.
1.
Highlight the PID option and press 'OK'.
2.
Select the 'Custom Calibration' icon.
3.
Zero calibration
4.
commence Zero calibration sequence. The display will countdown 30 seconds. Once
complete, select 'OK' to proceed. Disconnect the Carbon Filter Assembly and replace the
cap ends.
The useful life of the Carbon Filter Assembly will be shortened if it is open to
atmosphere for prolonged periods.