Chapter 6: Solving Problems
If you do not see the counter overflow indication "!" where you expected to see it in a trace list
If you do not see the counter overflow indication
"!" where you expected to see it in a trace list
This may be a normal indication. If you scroll through a reduced portion of the
trace memory, one that contains no counter overflow, no counter overflow
indication will be seen, even if counter overflow occurred before the line range you
specified in your display/store/copy command. The routine that reads trace
memory to compose a trace list only reads the portion of the trace memory you
specify in your display/store/copy command.
If the Performance Verification reports massive
pv failures
Check to make sure mapper chip U84 on Emulation Control Card Subassembly
HP 64748C is installed properly in its socket.
113