Agilent Technologies InfinityLab LC Series User Manual page 31

Variable wavelength detectors
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

2
Site Requirements and Specifications
Performance Specifications
Table 3
Feature
Flow cells
Spectral tools
Analog output
Instrument Control
Local Control
Communication
GLP
Safety and mainte-
nance
Agilent InfinityLab LC Series VWD User Manual
Performance Specifications 1260 Infinity II Variable Wavelength Detector (G7114A)
Specification
Standard:14 µL volume, 10 mm cell path length and 40 bar (588 psi) pres-
sure maximum
Micro:2 µL volume, 3 mm cell path length and 120 bar (1760 psi) pressure
maximum
Semi-micro:5 µL volume, 6 mm cell path length and 40 bar (588 psi) pres-
sure maximum
Preparative:4 µL volume, 3 mm cell path length and 120 bar (1760 psi) pres-
sure maximum
Preparative:0.3 mm cell path length and 50 bar (725 psi) pressure maximum
Preparative:0.06 mm cell path length and 50 bar (725 psi) pressure maxi-
mum
High pressure:14 µL volume, 10 mm cell path length and 400 bar (5801 psi)
pressure maximum
Stop-flow wavelength scan
Recorder/Integrator 100 mV or 1 V, 1 output
Lab Advisor B.02.08 or above
LC and CE Drivers A.02.14 or above
For details about supported software versions refer to the compatibility
matrix of your version of the LC and CE Drivers
Agilent Instant Pilot (G4208A) B.02.19 or above
Controller-area network (CAN), USB
ERI: ready, start, stop and shut-down signals
Early maintenance feedback (EMF) for continuous tracking of instrument
usage in terms of lamp burn time with user settable limits and feedback
messages. Electronic records of maintenance and errors. RFID for electron-
ics records of flow cell and UV lamp conditions (path length, volume, prod-
uct number, serial number, test passed, and usage). Verification of
wavelength accuracy with built-in holmium oxide filter.
Extensive diagnostics, error detection and display through Agilent Instant
Pilot and Agilent Lab Advisor software. Leak detection, safe leak handling,
leak output signal for shutdown of pumping system. Low voltages in major
maintenance areas. Tracking of flow cells and lamps with RFID (radio fre-
quency identification) tags
31