9
Maintenance
Replace the Deuterium Lamp
7
Insert the lamp (RFID tag on top) (1.) and fix the screws
(2.).
2.
9
Replace the fan cover (1.) and fix its screw (2.).
2.
Next Steps:
11 Reset the lamp counter as described in the User Interface documentation (required for non-RFID tag lamps only).
12 Turn the lamp ON and give the lamp more than 10 minutes to warm-up.
13 Perform a Wavelength Re-calibration after lamp warm-up.
After lamp on, the detector requires a warm-up time of 60 min. No
NOTE
measurements should be performed during this time.
Agilent InfinityLab LC Series VWD User Manual
1.
1.
8
Reconnect the connector.
10 Close the doors.
134