Fluke 5520A Programmer's Manual page 31

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

Command List (cont)
FUNC?
Response:
HARMONIC
Parameters: 1. Harmonic multiple
HARMONIC?
Response:
*IDN?
Responses:
INCR
Parameters: 1. Positive or negative step size
ISCE
Parameter:
(Character) DCV, ACV, DCI, ACI,
RES, CAP, RTD, TC_OUT,
DC_POWER, AC_POWER,
DCV_DCV, ACV_ACV, TC_MEAS,
or PRES_MEAS
2. (Optional) PRI (fundamental is at
primary output location) or SEC
(fundamental is at secondary
output location)
1. (Integer) Harmonic multiple
2. (Character) PRI or SEC (location
of fundamental)
(Indefinite ASCII) A message
containing four fields separated by
commas as follows:
1. Manufacturer
2. Model number
3. Serial number
4. Firmware revision levels for the
Main CPU, Front Panel CPU, and
Inguard PGA
2. (Optional) Magnitude
Decimal equivalent of the binary
number to load into the registers
29

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents