Olympus EVIS LUCERA TJF 260V Reprocessing Manual page 74

Duodenovideoscope
Table of Contents

Advertisement

Chapter 4 Cleaning and Disinfection Equipment
70
The elevator wire channel may not be properly high-level
disinfected by all reprocessors. If using a reprocessor, please
contact the manufacturer of the reprocessor for certification
that the elevator wire channel can be efficaciously high-level
disinfected in their unit. If the elevator wire channel cannot be
high-level disinfected by a reprocessor, clean and disinfect
the elevator wire channel according to procedures described
in Chapter 3, "Cleaning, Disinfection and Sterilization
Procedures".
If using the endoscope reprocessor, be sure to attach the
appropriate connectors (see Table 4.1). Otherwise,
insufficient cleaning and disinfection or sterilization of the
endoscope may pose an infection control risk to the patient
and/or operators performing the next procedure with the
endoscope.
For air/water
OER
EW-30
OER-A
Table 4.1
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
For instrument
channel, suction
channel
MH-861
MAJ-818
For elevator water
channel
MAJ-33
MAJ-819
channel
MAJ-41
MAJ-825

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Evis lucera jf 260v

Table of Contents