Olympus EVIS LUCERA TJF 260V Reprocessing Manual page 68

Duodenovideoscope
Table of Contents

Advertisement

Chapter 3 Cleaning, Disinfection and Sterilization Procedures
Manual cleaning
64
To prevent damage, make sure that the items immersed in
detergent solution do not contact one another.
1.
Fill a small basin with clean water. Use a basin which is deep enough to
allow the distal cover to be completely immersed.
2.
Fill a small basin with detergent solution at the temperature and
concentration recommended by the detergent manufacturer. Use a basin
which is deep enough to allow the distal cover to be completely immersed.
3.
Immerse the distal cover in the basin filled with detergent solution.
4.
Using the 30 cm
distal cover with detergent solution while the distal cover is immersed.
5.
Using a clean, soft brush or lint-free cloth, meticulously clean the external
surfaces in detergent solution.
6.
Cover the basin with a tight fitting lid to minimize the diffusion of detergent
vapors. Soak the distal cover at the temperature and concentration
recommended by the detergent manufacturer.
7.
Remove the distal cover from the detergent solution. If debris remains,
ultrasonically clean it at 33 – 48 kHz for 5 minutes.
8.
Immerse the distal cover in the clean water.
9.
Gently agitate the distal cover to thoroughly rinse them.
10.
Remove the distal cover from the water.
11.
Use a sterile lint-free cloth to thoroughly wipe and dry all external surface.
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
3
(30 ml) syringe, flush the interior and recessed parts of the