Olympus EVIS LUCERA TJF 260V Reprocessing Manual page 38

Duodenovideoscope
Table of Contents

Advertisement

Chapter 3 Cleaning, Disinfection and Sterilization Procedures
Preparation
Cleaning the external surface
34
To prevent damage to the endoscope, never immerse it
together with objects other than the equipment listed above.
1.
Fill a basin with detergent solution at the temperature and concentration
recommended by the detergent manufacturer. Use a basin which is at least
40 cm by 40 cm (16" by 16") in size and deep enough to allow the
endoscope to be completely immersed.
2.
Fill a basin with clean water. Use a basin which is at least 40 cm by 40 cm
(16" by 16") in size and deep enough to allow the endoscope to be
completely immersed.
1.
Immerse the endoscope in the detergent solution.
2.
With the endoscope immersed, use a soft brush or lint-free cloth to
thoroughly brush or wipe all external surfaces of the endoscope. Pay
particular attention to the air/water nozzle opening and the objective lens,
and ensure that all surfaces of the distal end are cleaned thoroughly (see
Figure 3.19).
Lint-free cloth
Figure 3.19
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
Air/water nozzle
opening
Objective lens