Olympus EVIS LUCERA TJF 260V Reprocessing Manual page 29

Duodenovideoscope
Table of Contents

Advertisement

5.
Immerse the distal end of the insertion tube in the clean water (see Figure
3.12).
6.
Switch the airflow regulator to maximum output ("HIGH" or "3").
7.
Depress the AW channel cleaning adapter to feed water through the
channels for 30 seconds. Release the adapter to feed air through the
channels for 10 seconds or more (see Figure 3.12).
Figure 3.12
8.
Turn the light source OFF.
Flush detergent solution and air into the elevator wire channel
1.
Connect the washing tube to the elevator channel plug (see Figure 3.13).
Figure 3.13
2.
Immerse the distal end of the insertion tube in the water.
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
AW channel
cleaning adapter
Clean water
Elevator channel plug
Washing tube