Chapter 3 Cleaning, Disinfection and Sterilization Procedures
20
4.
Attach the 30 cm
mesh immersed in rinsing water, withdraw the syringe plunger and confirm
that rinsing water is drawn into the syringe. Depress the plunger and confirm
that rinsing water is emitted from the distal end of the suction channel tube.
Confirm that water is not emitted from the suction port.
 Inspection of the channel cleaning brush (BW-20T)
1.
Confirm that the brush head and the metal tip of the distal end are securely
in place. Check for loose or missing bristles (see Figure 3.5 on page 16).
2.
Check for bends, scratches and other damage to the shaft.
3.
Check for debris on the shaft and/or in the bristles of the brush.
 Inspection of the channel-opening cleaning brush
(MH-507)
1.
Check for loose or missing bristles (see Figure 3.6 on page 17).
2.
Check for bends, scratches and other damage to the shaft.
3.
Check for debris on the shaft and/or in the bristles of the brush.
 Inspection of the washing tube (MH-974)
Check for cracks, scratches, flaws, debris and other damage (see Figure 3.7 on
page 17).
 Inspection of the suction cleaning adapter (MH-856)
Check for cracks, scratches, flaws, debris and other damage (see Figure 3.8 on
page 18).
 Inspection of the AW channel cleaning adapter (MH-948)
Check for cracks, scratches, flaws, debris and other damage (see Figure 3.9 on
page 18).