Chapter 3 Cleaning, Disinfection and Sterilization Procedures
Flushing detergent solution into the elevator wire channel
44
1.
Connect the washing tube to the elevator channel plug (see Figure 3.31).
Figure 3.31
2.
3
Using the 5 cm
into the elevator wire channel via the washing tube (see Figure 3.32).
5 cm3 (5 ml) syringe
Elevator wire channel
Endoscope connector
Figure 3.32
3.
Disconnect the washing tube from the endoscope and immerse it in
detergent solution.
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
Elevator channel plug
Washing tube
(5 ml) syringe, inject 15 cm
3
(15 ml) of detergent solution
Washing tube
Elevator channel
plug