Olympus EVIS LUCERA TJF 260V Reprocessing Manual page 27

Duodenovideoscope
Table of Contents

Advertisement

Aspirate detergent solution
1.
Turn the suction pump ON.
2.
Attach the biopsy valve cap.
3.
Immerse the distal end of the insertion tube in detergent solution. Depress
the suction valve and aspirate detergent solution into the instrument channel
for 30 seconds (see Figure 3.10).
4.
Remove the distal end of the insertion tube from the detergent solution and
immerse it in clean water. Depress the suction valve and aspirate clean
water into the instrument channel for 10 seconds.
5.
Remove the distal end of the insertion tube from the clean water. Depress
the suction valve and aspirate air for 10 seconds.
6.
Turn the suction pump OFF.
Figure 3.10
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
Monitor the suction bottle on the suction pump carefully to
ensure that it does not overflow. Otherwise, suction pump
damage could result.
Suction valve
Detergent solution
23