Olympus EVIS LUCERA TJF 260V Reprocessing Manual page 69

Duodenovideoscope
Table of Contents

Advertisement

High-level disinfection
1.
Fill a small basin with disinfectant solution at the temperature and
concentration recommended by the disinfectant manufacturer. Use a basin
which is deep enough to allow the distal cover to be completely immersed.
2.
Immerse the distal cover in the disinfectant solution.
3.
Using the 30 cm
distal cover with disinfectant solution while the distal cover is immersed.
4.
Using a clean lint-free cloth, wipe off any air bubbles adhering to surfaces.
5.
Cover the disinfectant basin with a tight fitting lid to minimize the diffusion of
disinfectant vapors. Soak the distal cover for the amount of time at the
temperature and concentration recommended by the disinfectant
manufacturer.
6.
Remove the distal cover from the disinfectant solution.
Rinse after high-level disinfection
After high-level disinfection, rinse the distal cover according to the procedures
described below.
 Sterile water rinse
1.
Fill a small basin with sterile water. Use a basin which is deep enough to
allow the distal cover to be completely immersed.
2.
Immerse the distal cover in the sterile water.
3.
Gently agitate the distal cover to thoroughly rinse.
4.
Remove the distal cover from the sterile water.
5.
Wipe and dry the distal cover with a sterile lint-free cloth.
6.
Dry the distal cover.
7.
Store the distal cover to following the instructions given in Chapter 5,
"Storage and Disposal".
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
3
(30 ml) syringe, flush the interior and recessed parts of the
Alcohol is flammable. Handle with care.
65

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Evis lucera jf 260v

Table of Contents