Olympus EVIS LUCERA TJF 260V Reprocessing Manual page 45

Duodenovideoscope
Table of Contents

Advertisement

Suction cleaning adapter
Endoscope connector
Figure 3.26
Flushing the interior of the forceps elevator
1.
Immerse the distal end in detergent solution, and operate the elevator
control lever to raise and lower the forceps elevator 3 times.
2.
With the forceps elevator raised, flush the interior of the forceps elevator
with detergent solution using the 30 cm
Figure 3.27
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
Weighted end
Insertion tube
Suction bottle
Forceps elevator
Instrument
channel port
Instrument
channel
Suction connector
Suction tube
Suction pump
3
(30 ml) syringe (see Figure 3.27).
Detergent solution
Suction channel
Suction
cylinder
Universal cord
41

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Evis lucera jf 260v

Table of Contents