Olympus EVIS LUCERA TJF 260V Reprocessing Manual page 46

Duodenovideoscope
Table of Contents

Advertisement

Chapter 3 Cleaning, Disinfection and Sterilization Procedures
Flushing detergent solution into the air/water channels
42
1.
Attach the channel plug's biopsy valve cap to the instrument channel port
(see Figure 3.28).
2.
Lower the channel plug's cylinder plug onto the air/water and suction
cylinders (see Figure 3.28).
3.
Push the plug down on the endoscope's control section and slide the plug
until it stops (see Figure 3.28).
Cylinder plug
Biopsy valve cap
Instrument
channel port
Figure 3.28
4.
Attach the injection tube's connector plug to the air and water supply
connectors on the endoscope connector (see Figures 3.29 and 3.30).
5.
Attach the injection tube's air pipe port to the air pipe on the endoscope
connector (see Figures 3.29 and 3.30).
6.
Attach the injection tube's suction channel tube to the suction connector on
the endoscope connector (see Figures 3.29 and 3.30).
7.
Immerse the suction port of the injection tube in the detergent solution.
8.
Attach the 30 cm
port (see Figure 3.30).
9.
3
Inject 90 cm
(90 ml) of detergent solution into the air/water channel.
10.
Detach the channel plug and injection tube from the endoscope, and leave
all items immersed.
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
Push
3
(30 ml) syringe to the injection tube's air/water channel
Air/water cylinder
Suction cylinder
Slide