Olympus EVIS LUCERA TJF 260V Reprocessing Manual page 19

Duodenovideoscope
Table of Contents

Advertisement

Reprocessing equipment parts and functions
For inspection of other equipment than that mentioned below, refer to the
instruction manual for the equipment being used.
 Water-resistant cap (MH-553)
The water-resistant cap is attached to the electrical connector on the endoscope
to protect the connector from water penetration during reprocessing. For leakage
testing, the venting connector on the water-resistant cap must be connected to
the leakage tester (see Figure 3.2).
Seal
Figure 3.2
 Channel plug (MH-944)
The channel plug is used to plug the openings of the instrument channel port,
air/water and suction cylinders during cleaning (see Figure 3.3).
Figure 3.3
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
Groove
Biopsy valve cap
Cylinder plug
Venting connector
15