Olympus EVIS LUCERA TJF 260V Reprocessing Manual page 50

Duodenovideoscope
Table of Contents

Advertisement

Chapter 3 Cleaning, Disinfection and Sterilization Procedures
Suction channel
Injection tube
Air/water channel port
Suction channel port
Dry external surfaces
46
8.
Using the 30 cm
3
with 90 cm
(90 ml) of air, respectively, via the injection tube (see Figures
3.30 and 3.33).
9.
3
Using the 5 cm
(10 ml) of air via the washing tube (see Figures 3.31 and 3.32).
10.
Remove the clean, lint-free cloth from the distal end and the control section
of the endoscope.
11.
Detach the channel plug and injection tube from the endoscope.
12.
Disconnect the washing tube from the endoscope.
Biopsy valve cap
Figure 3.33
1.
Using a clean lint-free cloth, wipe and dry the external surfaces of the
endoscope and all equipment thoroughly.
2.
Inspect the endoscope and all equipment for residual debris. Should debris
remain, repeat the procedures given in this section.
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
3
(30 ml) syringe, flush the air/water and suction channels
(5 ml) syringe, flush the elevator wire channel with 10 cm
Syringe
Channel plug
Suction cylinder
Suction channel
Suction connector
Suction port
3