Chapter 3 Cleaning, Disinfection and Sterilization Procedures
Removing the endoscope and all equipment from disinfectant
solution
50
4.
Soak the endoscope and all equipment in disinfectant solution for the
amount of time and at the temperature and concentration recommended by
the disinfectant manufacturer.
1.
Before removing the endoscope from the disinfectant solution, connect the
channel plug and injection tube to the endoscope. Attach the channel plug's
biopsy valve cap to the instrument channel port.
2.
Before removing the endoscope from the disinfectant solution, connect the
washing tube to the elevator channel plug.
3.
Remove the injection tube's suction port from the disinfectant solution.
4.
Using the 30 cm
air/water and suction channels respectively via the injection tube.
5.
3
Using the 5 cm
(10 ml) of air via the washing tube.
6.
Remove the endoscope and all equipment from the disinfectant solution.
7.
Disconnect all equipment from the endoscope.
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
3
(30 ml) syringe, flush 90 cm
(5 ml) syringe, flush the elevator wire channel with 10 cm
3
(90 ml) of air through the
3