Troubleshooting
Table 12-5
Status alarms and remedies continued
Alarm
code
Communicator
A005
Input Over-Range
A006
Transmitter Not
Characterized
A008
Density outside limits
A009
Transmitter
Initializing/Warming Up
A010
Calibration Failure
A011
Excess calibration
correction, zero too low
A012
Excess calibration
correction, zero too high
A013
Process too noisy to
perform auto zero
A014
Transmitter Failed
Configuration and Use Manual
ProLink II software
Possible remedy
Input Overrange
Check the test points. See Section 12.25.
Check the sensor coils. See Section 12.27.
Verify process.
Make sure that the appropriate measurement unit is
configured. See Section 12.20.
Verify 4 mA and 20 mA values. See Section 12.21.
Verify calibration factors in transmitter configuration.
See Section 6.2.
Re-zero the transmitter.
Not Configured
Check the characterization. Specifically, verify the
FCF and K1 values. See Section 6.2.
If the problem persists, contact Micro Motion. See
Section 12.3.
Density Overrange
Check the test points. See Section 12.25.
Check the sensor coils. See Section 12.27.
Verify process. Check for air in the flow tubes, tubes
not filled, foreign material in tubes, or coating in
tubes.
Verify calibration factors in transmitter configuration.
See Section 6.2.
Perform density calibration. See Section 10.5.
Transmitter
Allow the meter to warm up. The error should
Initializing/Warming Up
disappear once the meter is ready for normal
operation.
If alarm does not clear, make sure that the sensor is
completely full or completely empty. Verify sensor
configuration and wiring to sensor.
Calibration Failure
If alarm appears during a transmitter zero, ensure
that there is no flow through the sensor, then retry.
Cycle power to the meter, then retry.
Zero too Low
Ensure that there is no flow through the sensor, then
retry.
Cycle power to the meter, then retry.
Zero too High
Ensure that there is no flow through the sensor, then
retry.
Cycle power to the meter, then retry.
Zero too Noisy
Remove or reduce sources of electromechanical
noise, then attempt the calibration or zero procedure
again.
Sources of noise include:
• Mechanical pumps
• Pipe stress at sensor
• Electrical interference
• Vibration effects from nearby machinery
Cycle power to the meter, then retry. See
Section 12.24.
Transmitter Failed
Cycle power to the meter.
The transmitter might need service. Contact Micro
Motion. See Section 12.3.