Troubleshooting
Table 12-5
Status alarms and remedies continued
Alarm
code
Communicator
A027
Security Breach
A028
Sensor/Xmtr
Communication Failure
(2)
A031
Undefined
(3)
A032
Meter Verification Fault
Alarm
(4)
A032
Outputs Fixed during
Meter Verification
(2)
A033
Sensor OK / Tubes
Stopped by Process
(4)
A034
Meter Verification Failed
(4)
A035
Meter Verification
Aborted
A100
Primary mA Output
Saturated
A101
Primary mA Output
Fixed
A102
Drive Over-Range /
Partially Full Tube
A103
Data Loss Possible
A104
Calibration-In-Progress
A105
Slug Flow
A106
Burst Mode Enabled
A107
Power Reset Occurred
Configuration and Use Manual
ProLink II software
Possible remedy
Security Breach
Weights and Measures security seal has been
broken. Alarm can be cleared by user, but
authorized procedure is required to reestablish
security. See Chapter 11.
Core Processor Write
Cycle power to the meter.
Failure
The transmitter might need service or upgrading.
Contact Micro Motion. See Section 12.3.
Low Power
The core processor is not receiving enough power.
Check the power supply to the transmitter, and
check power wiring between the transmitter and the
core processor (4-wire remote installations only).
Meter
Meter verification in progress, with outputs set to
Verification/Outputs In
fault. Allow the procedure to complete.
Fault
If desired, abort the procedure and restart with
outputs set to last measured value.
Meter Verification In
Meter verification in progress, with outputs set to
Progress and Outputs
Fault or Last Measured Value. Allow the procedure
Fixed
to complete.
If desired, abort the procedure and restart with
outputs set to Continue Measurement.
Sensor OK / Tubes
No signal from LPO or RPO, suggesting that sensor
Stopped by Process
tubes are not vibrating. Verify process. Check for air
in the flow tubes, tubes not filled, foreign material in
tubes, or coating in tubes.
Meter Verification Failed
Rerun the test. If the test fails again, see
Section 10.3.4.
Meter Verification
If desired, read the abort code, see Section 10.3.4,
Aborted
and perform the appropriate action.
Primary mA Output
See Section 12.18.
Saturated
Primary mA Output
Check the HART polling address. See
Fixed
Section 12.19.
Exit mA output trim. See Section 5.4.
Exit mA output loop test. See Section 5.3.
Check to see if the output has been fixed via digital
communication.
Drive Overrange
Excessive drive gain. See Section 12.25.3.
Check the sensor coils. See Section 12.27.
Data Loss Possible (Tot
Cycle power to the meter.
and Inv)
View the entire current configuration to determine
what data were lost. Configure any settings with
missing or incorrect data.
The transmitter might need service. Contact Micro
Motion. See Section 12.3.
Calibration in Progress
Allow the meter to complete calibration.
Slug Flow
See Section 12.17.
Burst Mode Enabled
No action required.
Power Reset Occurred
No action required.