Troubleshooting
Table 12-5
Status alarms and remedies continued
Alarm
code
Communicator
(5)
A108
Event #1 Triggered
(5)
A109
Event #2 Triggered
A110
Frequency Output
Saturated
A111
Frequency Output Fixed
(6)
A112
Software upgrade
recommended
A113
Secondary mA Output
Saturated
A114
Secondary mA Output
Fixed
A115
External Input Error
A116
API Temperature
Out-of-Limits
A117
API Density
Out-of-Limits
A118
Discrete Output 1 Fixed
A119
Discrete Output 2 Fixed
A120
ED: Unable to Fit Curve
Data
A121
ED: Extrapolation Alarm
(3)
A131
Meter Verification Info
Alarm
(4)
A131
Meter Verification in
Progress
(2)
A132
Simulation Mode Active
140
ProLink II software
Event 1 Triggered
Event 2 Triggered
Frequency Output
Saturated
Frequency Output Fixed
S/W Upgrade
Recommended
Secondary mA Output
Saturated
Secondary mA Output
Fixed
External Input Error
API: Temperature
Outside Standard
Range
API: Density Outside
Standard Range
Discrete Output 1 Fixed
Discrete Output 2 Fixed
ED: Unable to Fit Curve
Data
ED: Extrapolation Alarm
Meter
Verification/Outputs at
Last Value
Meter Verification In
Progress
Simulation Mode Active
Possible remedy
Be advised of alarm condition.
If you believe the event has been triggered
erroneously, verify the Event 1 settings. See
Section 8.11.
Be advised of alarm condition.
If you believe the event has been triggered
erroneously, verify the Event 2 settings. See
Section 8.11.
See Section 12.18.
Exit frequency output loop test.
Contact Micro Motion to get a Series 1000/2000
transmitter software upgrade. See Section 12.3.
Note that the device is still functional.
See Section 12.18.
Exit mA output loop test. See Section 5.3.
Exit mA output trim. See Section 5.4.
Check to see if the output has been fixed via digital
communication.
HART polling connection to external device has
failed. Ensure that external device is available:
• Verify device operation.
• Verify wiring.
Verify polling configuration. See Section 9.4.
Verify process.
Verify API reference table and temperature
configuration. See Section 8.6.
Verify process.
Verify API reference table and density configuration.
See Section 8.6.
Exit discrete output loop test. See Section 5.3.
Exit discrete output loop test. See Section 5.3.
Verify enhanced density configuration.
Verify process temperature.
Verify process density.
Verify enhanced density configuration.
Meter verification in progress, with outputs set to
last measured value. Allow the procedure to
complete. If desired, abort the procedure and restart
with outputs set to fault.
Meter verification in progress, with outputs set to
continue reporting process data. Allow the
procedure to complete.
Disable simulation mode. See Section 12.10.