Connecting with ProLink II or Pocket ProLink Software
5. Start ProLink II. Choose
specify connection parameters appropriate to your connection type:
Service port mode – Immediately after the transmitter is powered up, terminals 33 and 34
are available in service port mode for 10 seconds. To connect during this period, set
Protocol
rate
,
Table 3-2). If a connection is made during this period, the port will remain in service port
mode until power is cycled.
RS-485 mode – If no connection is made during the 10-second period, the terminals are
automatically reset to the configured RS-485 communication parameters. To connect, set
the connection parameters to the values configured in your transmitter (see Table 3-2).
Table 3-2
Modbus connection parameters for ProLink II
Connection parameter
Protocol
Baud rate
Stop bits
Parity
Address/Tag
COM port
(1) Required value; cannot be changed by user.
6. Click
Connect
7. If an error message appears:
a. Swap the leads between the two terminals and try again.
b. Ensure you are using the correct COM port.
c. If you are in RS-485 mode, you may be using incorrect connection parameters.
Connect using the service port and check the RS-485 configuration. If required,
change the configuration or change your RS-485 connection parameters to match the
existing configuration.
If you are unsure of the transmitter's address. use the
window to return a list of all devices on the network.
Check all the wiring between the PC and the transmitter.
Configuration and Use Manual
Connection > Connect to Device
to Service Port, and set
Stop bits
, and
Parity
are set to standard values and cannot be changed (see
Configurable (RS-485 mode)
As configured in transmitter
(default = Modbus RTU)
As configured in transmitter (default = 9600)
As configured in transmitter (default = 1)
As configured in transmitter (default = odd)
Configured Modbus address (default = 1)
COM port assigned to PC serial port
.
. In the screen that appears,
to the appropriate value for your PC.
COM port
Connection type
SP standard (service port mode)
Modbus RTU
(1)
38,400
(1)
1
(1)
none
(1)
111
COM port assigned to PC serial port
Poll
Baud
(1)
button in the
Connect
25