Measurement Performance
In addition, if your transmitter has a discrete input, you can configure the discrete input to initiate a
Smart Meter Verification test remotely.
You can use these methods in any combination. For example, you can specify that a Smart Meter
Verification test will be executed three hours from now, every 24 hours starting now, every time a
specific discrete event occurs, and every time the discrete input is activated.
To define meter verification as an event action, see Section 6.8
To define meter verification as a discrete input action, see Section 6.8
To set up a one-time automatic execution, set up a recurring execution, view the number of
hours until the next scheduled test, or delete a schedule:
-
With ProLink II, click
-
With the display, see Figures 10-5 and 10-11.
-
With the Communicator, see Figure 10-12.
Note the following:
If you are setting up a one-time automatic execution, specify the start time as a number of
hours from the present time. For example, if the present time is 2:00 and you specify 3.5 hours,
the test will be initiated at 5:30.
If you are setting up a recurring execution, specify the number of hours to elapse between
executions. The first test will be initiated when the specified number of hours has elapsed, and
testing will be repeated at the same interval until the schedule is deleted. For example, if the
present time is 2:00 and you specify 2 hours, the first test will be initiated at 4:00, the next at
6:00, and so on.
If you delete the schedule, both the one-time execution and the recurring execution settings are
deleted.
116
Tools > Meter Verification > Schedule Meter Verification
Micro Motion Series 1000 and Series 2000 Transmitters
.