3/Maintenance

CLEANING

Do not use aromatic hydrocarbons (such as gasoline or
other fuels) or chlorinated solvents for cleaning. They
may damage plastic materials used in the instrument.
Do not use detergent of any kind for cleaning the PCA.
Clean the instrument case with a mild detergent and water.
The main and top PCAs may be washed with isopropyl alcohol or deionized
water and a soft brush. Dry with clean dry air at low pressure.

VECTOR OUTPUT I/O MODULE SELF TEST

The Vector Output I/O Module Self Test verifies that the module is
connected and is communicating with the system. Use the following
procedure to self test the module:
1.
Press MAIN MENU on the mainframe and move the cursor to the first
(left-most) field using the left arrow ( <- ) key.
2.
Press the SELFTEST softkey (F1).
3.
Move the cursor one field to the right and press the I/O MOD softkey
(F3).
4.
Move the cursor one more field to the right, and press the number of
the I/O Module to be tested. Check that the display reads:
MAIN: SELFTEST I/O MOD <n>
(where <n> signifies the number of the I/O Module).
5.
Press ENTER to initiate the self test.
If the self test fails, a failure message is displayed. Check the Vector
Output I/O Module connection and repeat the test.
If the test fails a second time, the Vector Output I/O Module requires
service.

DISASSEMBLY

To gain access to the 2 PCAs in the Vector Output I/O Module, perform
the following steps:
1.
Disconnect the module from the mainframe.
2.
Turn the module over on its top (with the large module decal facing
up). Remove the four Phillips screws that hold the case halves
together, and separate the top and bottom case halves.
CAUTION
3-2