5. After all files have been copied, remove the floppy disk from the analyzer's disk drive
and exit both copies of Windows Explorer.
Set Up the Agilent Administrator Password
If the hard disk has been replaced, it is necessary to set up the Agilent administrator
password for the analyzer. This requires obtaining a password from Agilent:
Obtaining a Password
• If this procedure is being performed by Agilent personnel, go to the following Internet
site and follow the instructions under "Agilent Administrator Password".
http://mktwww.soco.agilent.com/field/service/network/pna/support.htm
• If this procedure is being performed by non-Agilent personnel, contact Agilent for a
password before beginning this procedure. Refer to
Entering Your Agilent Administrator Password
The network analyzer must be powered up and operating for this procedure.
1. Exit the Network Analyzer program, if it is running.
2. On the Windows Desktop, right-click on My Computer, and then click Manage.
3. In the Computer Management window, navigate to System Tools, Local Users and
Groups, Users.
4. Right-click on Agilent, and then click on Set Password.
a. Type the password obtained from Agilent as the Password. Refer to
Password" on page 7-67
b. Type the password again in the Confirm password block.
c. Click OK.
5. Right-click on Agilent, and then click on Properties.
a. Unselect User must change password at next logon (if not already unselected).
b. Unselect User cannot change password (if not already unselected).
c. Select Password never expires (if not already selected).
d. Click OK.
6. Close the Computer Management window.
Post-Repair Procedures
Perform the post-repair adjustments, verifications, and performance tests that pertain to
this removal procedure. Refer to
Service Guide E8364-90026
for information if you have not yet obtained this password.
Table on page
7-76.
Repair and Replacement Procedures
Removing and Replacing the A41 Hard Disk Drive
"Contacting Agilent" on page
"Obtaining a
2-10.