Agilent Technologies E8362B Service Manual page 136

Pna series microwave network analyzers
Hide thumbs Also See for E8362B:
Table of Contents

Advertisement

Troubleshooting
Measurement System Troubleshooting
5. If this signal is not present, replace the A9 fractional-N synthesizer board. Refer to
"Removing and Replacing the A6, A8, A9, and A10 Boards" on page
6. If the signal is present, leave the spectrum analyzer connected to A9J106 and continue
with
"Checking the A9 Fractional-N Synthesizer Bands 3 through 25 Output" on
page
4-54.
Checking the A9 Fractional-N Synthesizer Bands 3 through 25 Output
Perform this procedure if you observe a problem in bands 3 through 25 in all receivers.
1. Set the network analyzer to a CW frequency of 2 GHz.
• The spectrum analyzer should measure a signal at 2.008 GHz.
2. If the signal is not present, replace the A9 fractional-N synthesizer board. Refer to
"Removing and Replacing the A6, A8, A9, and A10 Boards" on page
3. If the signal is present, reconnect cable W91, and then continue with
Source 20 Output" on page
Checking the A12 Source 20 Output
Perform this procedure if you observe a problem in any band and all signals from the A9
fractional-N synthesizer are present.
1. Refer to the block diagram at the end of this chapter and to
Including 080" on page
10 output.
CAUTION
Be careful not to damage the center pins of the semirigid cable. Some flexing
of the cables is necessary to measure the output.
2. Using a 5/16-inch torque wrench, disconnect W93 at the output of the A12 source 10
board.
3. Connect the spectrum analyzer to the A12 source 10 board.
4. Set the network analyzer to a CW frequency of 1 GHz.
• The spectrum analyzer should measure a signal at 1 GHz.
5. If the signal is not present, replace the A12 source 20 board. Refer to
Replacing the A12 Source 20" on page
6. If the signal is present, reconnect cable W93, and then continue with
Frequency Offset Receiver Band 0 Output" on page
4-54
4-54.
6-18. Locate the semirigid cables W93 at the A12 source
7-22.
PNA Series Microwave Network Analyzers
E8362B, E8363B, E8364B
7-18.
7-18.
"Checking the A12
"Top Cables, All Options
"Removing and
"Checking the A13
4-55.
Service Guide E8364-90026