Agilent Technologies E8362B Service Manual page 77

Pna series microwave network analyzers
Hide thumbs Also See for E8362B:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
E8362B, E8363B, E8364B
Procedure
1. Connect the equipment as shown in
network analyzer and the power meter.
Figure 3-19 Setup 1 for the Receiver Calibration Adjustment
2. On the System menu, point to Service, Adjustments, and then click
Receiver Calibration.
3. Ensure the GPIB settings are correct.
4. Click Calibrate, and then follow the instructions as they are displayed.
Figure 3-20 Setup 2 for the Receiver Calibration Adjustment
Service Guide E8364-90026
Figure
3-19. Connect a GPIB cable between the
Tests and Adjustments
Adjustments
3- 45