Agilent Technologies E8362B Service Manual page 73

Pna series microwave network analyzers
Hide thumbs Also See for E8362B:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
E8362B, E8363B, E8364B
TEST RECORD FOR POWER METER ACCURACY TEST
Power Meter Tested: Model No.:
Date:
Range Calibrator: Model No.:
Digital Voltmeter: Model No.:
Range
Calibrator
DVM Reading
Setting
(Vdc)
1 mW
A =
300 µW
B =
100 µW
D =
R = B/A =
(C/10)
S = 10
300 µW
% ERROR = ((R-S)/R) × 100 =
Limits: ±0.13%
Pass
T = D/A =
(E/10)
U = 10
100 µW
% ERROR = ((T-U)/T) × 100 =
Limits: ±0.10%
Pass
Service Guide E8364-90026
Tested by:
Test Equipment Used
11683A
Test Results
Channel A
Power Meter
Reading (dBm)
0.000 ±0.001 A =
C =
E =
Pass/Fail Calculations
=
Fail
=
Fail
Performance Tests (Agilent N2721A Software Package)
Serial No:
Serial No.:
Serial No.:
Channel B
DVM Reading
(Vdc)
B =
D =
R = B/A =
(C/10)
=
S = 10
% % ERROR = ((R-S)/R) × 100 =
Limits: ±0.13%
Pass
T = D/A =
(E/10)
=
U = 10
% % ERROR = ((T-U)/T) × 100 =
Limits: ±0.10%
Pass
Tests and Adjustments
Power Meter
Reading (dBm)
0.000 ±0.001
C =
E =
%
Fail
%
Fail
3- 41