Power Level Accuracy Test

This test verifies the E4991A's test signal power level accuracy with a power meter and a
power sensor.
Specification
Power Level Accuracy:
Required Equipment
Power Meter
Power Sensor
Multimeter
BNC(m)-BNC(m) Cable, 61 cm
N(m)-BNC(f) Adapter
7mm-N(f) Adapter
Dual Banana-BNC(f) Adapter
Feed Through Termination
Procedure
1. Connect the equipment as shown in Figure 2-3.
NOTE
Refer to Page 42 for the test setup applied to Option 007 and, Page 44 for Option 010.
Chapter 2
2 dB @  1 GHz
+2 dB/-4 dB @  1 GHz (with Opt. 007)
5.5 dB @  1 GHz (with Opt. 010)
3 dB @ > 1 GHz
+3 dB/-6 dB @ > 1 GHz (with Opt. 007)
7.6 dB @ > 1 GHz (with Opt.010)
Description
E4418A/B, E4419A/B or 438A
E4412A, E9304A or 8482A+8481D
3458A
p/n 8120-1839
p/n 1250-0780
11524A
p/n 1251-2277
11048C
Performance Test and Function Test
Performance Test
Recommended Model
31