Repair and Replacement Procedures
Removing and Replacing the A41 Hard Disk Drive
Removal Procedure
1. Disconnect the power cord.
Removing the HDDA
Refer to
Figure 7-30
1. With a T-10 TORX driver, loosen the four HDDA mounting screws (item
panel.
2. Remove the HDDA from the analyzer by pulling on the finger grip (item
interconnects with a connector inside the analyzer rear panel so moderate force may be
necessary to disengage this connector.
Figure 7-30 Hard Disk Drive Assembly Removal
Disassembling the HDDA
When replacing the hard disk, first torque all four screws (item ③) to 4 in-lb,
IMPORTANT
then secure each to 7 in-lb with a T-10 TORX driver.
Refer to
Figure
7-31.
1. Remove the hard disk drive assembly from the analyzer. Refer to
on page
7-64.
2. With a T-10 TORX driver, remove the four screws (item
(item
), that mount the hard disk (item
3. Disconnect the hard disk (item
7-64
for this procedure.
) from the interface board (item
PNA Series Microwave Network Analyzers
), and the eight bushings
), to the bracket.
Service Guide E8364-90026
E8362B, E8363B, E8364B
) on the rear
). The HDDA
"Removing the HDDA"
).