Agilent Technologies E8362B Service Manual page 117

Pna series microwave network analyzers
Hide thumbs Also See for E8362B:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
E8362B, E8363B, E8364B
Node Descriptions, A10 Frequency Reference Board
Node 211
100 MHz Tune Node 211 senses the tune voltage that phase locks the
100 MHz VCO. The tune voltage is set by referencing the 100 MHz VCO to
the internal 10 MHz OCXO or to the 10 MHz EXT REF.
Node 212
100 MHz Level Node 212 senses the average voltage output level of the
100 MHz VCO phase locked reference signal.
Node 213
99.50 MHz Tune Node 213 senses the tune voltage that phase locks the
99.50 MHz VCO. The tune voltage is set by referencing the 100 MHz VCO.
Node 214
99.50 MHz Level Node 214 senses the average voltage output level of the
99.50 MHz reference VCO.
Node 215
External Reference Detect Node 215 senses a TTL signal which is used
to set the reference frequency switch. The node registers a low TTL voltage
when the internal 10 MHz OCXO signal is used or a high TTL voltage when
the external 10 MHz EXT REF is used.
Node 216
Phase Lock Average Voltage Node 216 senses the average voltage of the
8.333 MHz phase lock reference signal that is routed to the A11 phase lock
board. This voltage is constant through all bands.
Node 217
500 kHz Level Node 217 senses the average voltage of the 500 kHz phase
lock reference signal.
Node 218
33.1667 MHz 2nd LO Average Voltage Node 218 senses the average
voltage of the 33.1667 MHz 2nd LO signal that is routed to the A35 receiver
motherboard. This voltage is constant through all bands.
Node Descriptions, A11 Phase Lock Board
Node 311
NC Node 311 is not used in this product.
Node 312
100 kHz Phase Detector Node 312 senses the average voltage of the
phase comparator at the output of a 700 kHz low-pass filter.
Node 313
NC Node 313 is not used in this product.
Node 314
100 kHz Phase Detector Integrator Node 314 senses the tune voltage
that phase locks the source. The integrator that supplies this signal receives
its input from the 700 kHz phase detector (Node 312).
Node 315
1V/GHz YTO Drive Voltage Node 315 senses the YIG-tuned oscillator
drive voltage which sweeps at 1V/GHz.
Node 316
YTO Drive Current Node 316 senses the YIG-tuned oscillator drive
current.
Node 317
YTO Tune Compensation Node 317 senses the voltage output of the YTO
tune compensation summation node.
Node 318
Ground Node 318 is connected to digital ground to serve as a
measurement reference.
Service Guide E8364-90026
Troubleshooting
Measurement System Troubleshooting
4- 35

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8363bE8364b

Table of Contents