Tests and Adjustments
Adjustments

3.8 GHz PMYO Frequency Adjustment

The 3.8 GHz adjustment is used to adjust the frequency accuracy of the network analyzer's
3.8 GHz permanent-magnet YIG oscillator (PMYO) on the A12 source 20.
Equipment used for the Frequency Adjustment at 3.8 GHz
Equipment Type
Spectrum analyzer
RF Cable, 3.5 mm (f) to 3.5 mm (f)
Adapter, 3. 5 mm (m) to type-N (m)
RF Cable, 2.4 mm (f) to 2.4 mm (f)
Adapter, 2.4 mm (m) to type-N (m)
a. Must be capable of measuring a signal at 3.8 GHz.
Procedure
1. Connect the equipment as shown in
network analyzer and the spectrum analyzer.
Figure 3-24 Setup for Adjustment of the 3.8 GHz PMYO
2. On the System menu, point to Service, Adjustments, and then click
3.8 GHz Freq Adjust.
3. Ensure the GPIB settings are correct.
4. Click Begin Adj, and then follow the instructions as they are displayed.
3-50
Models
Model or
Used With
Part Number
All
8565E
E8362B
85131C
E8362B
1250-1743
E8363B, E8364B
85133C
E8363B, E8364B
11903A
Figure
3-24. Connect a GPIB cable between the
PNA Series Microwave Network Analyzers
E8362B, E8363B, E8364B
Alternate Model
or Part Number
a
856xE
85131E
None
85133E
None
Service Guide E8364-90026