Agilent Technologies E8362B Service Manual page 126

Pna series microwave network analyzers
Hide thumbs Also See for E8362B:
Table of Contents

Advertisement

Troubleshooting
Measurement System Troubleshooting
Table 4-8
LO Path Frequencies
Network Analyzer
Model
Band
Frequency Band
of Problem (GHz)
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
4-44
Network Analyzer
Source Frequency
Setting (GHz)
0.010 to 0.045
0.045 to 0.748
0.748 to 1.500
1.50 to 3.00
3.00 to 3.80
3.80 to 4.50
4.50 to 4.80
4.80 to 6.00
6.00 to 6.40
6.40 to 7.60
7.60 to 10.00
10.00 to 12.00
12.00 to 12.80
12.80 to 15.20
15.20 to 16.00
16.00 to 20.00
20.00 to 22.80
22.80 to 25.60
25.60 to 30.00
30.00 to 32.00
32.00 to 36.00
36.00 to 38.40
38.40 to 40.00
40.00 to 45.60
45.60 to 48.00
48.00 to 50.00
PNA Series Microwave Network Analyzers
Expected A17
Output
Frequency
(GHz)
0.025
0.026
0.500
0.508
1.000
1.008
2.200
2.208
3.400
3.408
4.200
4.208
4.600
4.608
5.400
5.408
6.200
6.208
7.000
7.008
8.800
8.808
11.000
5.504
12.400
6.204
14.000
7.004
15.600
7.804
18.000
9.004
21.400
7.136
24.200
8.069
27.800
9.269
31.000
5.168
34.000
5.668
37.200
6.201
39.200
6.535
42.800
7.135
46.800
7.801
49.000
8.168
Service Guide E8364-90026
E8362B, E8363B, E8364B
Expected A18
Output
Frequency
(GHz)
0.026
0.508
1.008
2.208
3.408
4.208
4.608
5.408
6.208
7.008
8.808
11.008
12.408
14.008
15.608
18.008
7.136
8.069
9.269
10.336
11.336
12.403
13.069
14.269
15.602
16.336

Advertisement