PNA Series Microwave Network Analyzers
Theory of Operation
E8362B, E8363B, E8364B
Signal Separation Group Operation
Normal Option 014 Configuration
The normal configuration for the Option 014 is shown in
Figure 5-4 on page
5-19. With this
configuration and inclusion of an external amplifier and accessories, you can calibrate the
analyzer and test devices at power levels up to +30 dBm. You can make measurements in
the forward, reverse or both directions and still achieve these high power levels. For more
information on higher power measurements, search for "Option 014" in the embedded help
index in the analyzer.
Figure 5-4 Signal Separation Group, Option 014 Analyzer
Service Guide E8364-90026
5- 19