D
Select Negotiation Details to see details.
On the Negotiation Details screen, Yes for Pin Reversal
indicates a reversed pair on the link, (such as wires 1 and 2
crossed). Run a wire map test on the link to check the
wiring.
E
Select DTX Address to see details, including the subnet
mask.
F
Network addresses used during the test:
Gateway: Address of the network device that joins
networks using different protocols or passes data
between networks.
DHCP Server: Dynamic Host Configuration Protocol
server address. Shown only in DHCP mode.
DNS Server: Domain Name System server address.
In DHCP mode, the addresses above show as 0.0.0.0 if
the DHCP server does not provide them.
Figure 24. Network Connectivity Results Screen (DHCP example for twisted pair) (cont.)
To see ping results for the above devices, highlight a
H
device; then press
The checkmarks and Xs indicate how many replies the
tester received to its three pings:
: Green: Ping replies received for all requests.
: Orange: At least one ping reply received.
: No ping replies received, indicating a problem
with the connection.
G
<Target Address name>: appears if you use L Ping
to ping a device. See page 48.
H
See "Monitoring Network Traffic" on page 50 and
"Blinking a Port Light" on page 50.
I
L
Press
Ping to ping devices. See page 48.
Verifying Network Service
. See Figure 25.
47