Theoretical lateral resolution:
Wavelength = emission, in nanometers, usually the central wavelength of the emission band of the
barrier filter.
NA = numerical aperture of the objective lens.
Examples:
Using a 20X/NA0.75 objective with the green channel, 540 nm/2X.75 = 540 nm/1.5 = 360 nm lateral
resolution (0.36 µm).
Zoom factor for 2X oversampling =
12.7 Manual shutter
(open position)
12.8 Current mirror position
12.1 Focus Knob

Figure 12. DIC and Focus Controls, Lower Half of the Microscope

May 11, 2011
Olympus Fluoview-1000 User's Guide
wavelength(nm)
2(NA)
Pixel size
(wavelength)(NA)
40
(2)
12.6 DIC Analyzer
Adjustment knob
12.5 DIC Analyzer
Locking screw
12.4 DIC Analyzer slider
(out position)
12.3 Lens
Drop/Raise
12.2 F/C Coarse/Fine Focus