play in the center-pivot ball, such that the turret is not
held in the exact center point, the whole turret may
move slightly, resulting in poor objective centering and
all that that entails. This problem may be encountered
if the threaded retaining ring has not been properly
snugged down, or if the center pivot-adjustment screw
has not been properly adjusted to remove the play in
the center-pivot mechanism.