Replacing HFII Assembly Wire Harness

NOTE: For this procedure, refer to FIGURE 8.4 on page 46.
1. Remove the cabinet. Refer to Removing/Installing the Cabinet on page 21.
2. Remove the control panel. Refer to Removing/Installing Control Panel on page 25.
3. Note the orientation of the connectors on the power inlet (Detail "A").
4. Disconnect the two connectors from the power inlet.
5. Note the orientation of the two connectors on the hour meter (Detail "A").
6. Disconnect the two connectors from the hour meter.
Part No 1141487
Power Inlet
DETAIL "C"
FIGURE 8.3 Replacing Bleed Resistor Assemblies
SECTION 8—WIRING ASSEMBLIES
DETAIL "B"
Motor Assembly
Grounding Clip
Bleed Resistor
45
Bleed Resistor
Assembly 2
Assembly 1
HomeFill™II Compressor