Testing/Replacing The Power Switch - Invacare HomeFill II Service Manual

Hide thumbs Also See for HomeFill II:
Table of Contents

Advertisement

Testing/Replacing the Power Switch

NOTE: For this procedure, refer to FIGURE 6.3 on page 27.
Tools required for this procedure are:
• Ohmmeter
1. Remove the cabinet. Refer to Removing/Installing the Cabinet on page 21.
2. Remove the control panel. Refer to Removing/Installing Control Panel on page 25.
3. Note the location of the two wire connectors of the power switch for later replacement.
4. Remove the two wire connectors from the power switch.
A. Using an ohmmeter, test the continuity of the switch by placing the meter leads on
the switch terminals where the wires were connected.
B. Turn the switch to the On position.
C. If no indication of continuity is achieved, proceed to step 5. If continuity is
achieved, proceed to Step 7.
5. Release the four retaining clips and push the power switch out of the control panel.
6. Install power switch into control panel as shown.
7. Install the two wire connectors onto the power switch.
8. Install the control panel. Refer to Removing/Installing Control Panel on page 25.
9. Install the cabinet. Refer to Removing/Installing the Cabinet on page 21
Retaining Clips
Control Panel
Power Switch
Retaining Clips
Part No 1141487
FIGURE 6.3 Testing/Replacing the Power Switch
SECTION 6—PANEL ASSEMBLY
Wire
Connectors
27
HomeFill™II Compressor

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents