SECTION 9—TROUBLESHOOTING/CYLINDER FILL TIMES
PROBLEM
RED Light On
Compressor cycles On
and Off
or
O
BELOW NORMAL
2
(YELLOW) and FILLING
(GREEN) lights are alter-
nating
or
O
BELOW NORMAL
2
(YELLOW) light goes On
and Off
O
BELOW NORMAL
2
(YELLOW) light stays On.
Fill times exceed
maximum as
recommended fill times or
GREEN
FILLING light stays On.
Excessive noise when
compressor is On.
Hissing sound from Regu-
lator when flow selector
is Off.
Regulator is On but O
is
2
not being delivered.
HomeFill™II Compressor
1.Check the cylinder gauge to see if it is FULL. Cylinder full - Remove the cylinder and
turn the unit Off. Perform the Cylinder Prefill Inspection on page 14 with a cylinder
that is NOT FULL.
Cylinder not full- Proceed to STEP 2. Cylinder MUST be below 1500 psi for
compressor to start. If the psi is above 1500 the compressor identifies the cylinder as
full.
2. Check that the cylinder is connected properly. Turn the compressor Off. Remove
the cylinder and reinstall. Turn the compressor On.
3. If cylinder is not full and is connected properly, the compressor may have internal
failure. Contact Invacare for service.
1. Check the concentrator liter flow setting is at 3 L/min. or less.
If necessary, adjust the liter flow setting (refer to the HomeFill II owner's manual).
Wait at least 20 minutes for the GREEN light to appear. If it does not appear, proceed
to STEP 2.
2. The concentrator may need service, contact Invacare for service.
1. Ensure the concentrator has warmed up for at least 30 minutes. If it has not been
30 minutes, turn the compressor Off until the concentrator warms up. If the
YELLOW light still remains on proceed to STEP 2.
2. Inspect the connection between the compressor and the concentrator for damage
or pinching. If damaged, replace. If pinched, straighten tubing. Turn the compressor
Off for at least 10 seconds.
Turn the compressor On. If the YELLOW light still remains On proceed to STEP 3.
3. The concentrator may need service, contact Invacare for service.
1. Ensure the cylinder regulator is set to Off.
2. Check the recommended fill times shown in Cylinder Fill Times on page 49.
3. Check that the cylinder is connected properly. Turn the compressor Off. Remove
the cylinder and reinstall. Turn the compressor On.
4. Check the compressor/concentrator line for proper connection and/or damage
(i.e., leaks or kinks).
5. Check the concentrator liter flow setting is at three L/min. or less. If necessary,
adjust the liter flow setting (refer to the HomeFill II owner's manual). If the GREEN
filling light does not go Off, the compressor and/or concentrator may need service.
Contact Invacare for service.
1. Contact Invacare for service.
1. Turn flow selector Off.
2. If hissing sounds continue, contact Invacare for service.
1. Refill cylinder.
2. Check that cannula is straight and not pinched. If damaged, replace.
3. Contact Invacare for service.
SOLUTION
48
Part No 1141487