SECTION 7—BASE COMPONENTS
8. Connect the four wire connectors to the transformer in the orientation (for colors and
sizes) noted previously.
9. Install the cabinet. Refer to Removing/Installing the Cabinet on page 21.
Mounting Screw
Wire Connector

Testing/Replacing Capacitor

Testing

The capacitor is an electrical storage device that stores an electronic charge.
ALWAYS discharge the capacitor before testing or handling. Failure to discharge
the capacitor may result in bodily harm and/or damage to test equipment.
NOTE: For this procedure, refer to FIGURE 7.6 on page 37.
1. Remove the cabinet. Refer to Removing/Installing the Cabinet on page 21.
2. Remove the two wire connectors from the capacitor.
3. Discharge the capacitor by placing a screw driver with an insulated handle across the
wire connectors on the top of the capacitor.
4. Connect the leads of a capacitance meter to the terminals on the capacitor.
NOTE: For the capacitor to be within specification, the meter should read 10 mfd +10% or –5%.
5. Perform one of the following:
• If the capacitor is out of specification, replace the capacitor. Refer to Replacing
Capacitor on page 37.
• If the capacitor is in specification, reconnect the two wires and proceed to STEP 7.
6. Install the cabinet. Refer to Removing/Installing the Cabinet on page 21.
HomeFill™II Compressor
FIGURE 7.5 Replacing Transformer
WARNING
36
Transformer
Mounting Screw
Base
Wire Connector
Part No 1141487

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents