SECTION 7—BASE COMPONENTS
Removing Connector Fillport Dust Cover and Lanyard -
Units Manufactured Before 10/03/2005
DANGER
To prevent electrical shock, ALWAYS disconnect compressor from electrical
outlet before servicing.

Removing

NOTE: For this procedure, refer to FIGURE 7.7 on page 39.
Tools required for this procedure are:
• 7/16 Wrench, Phillips Screwdriver
NOTE: Because the parts for the this style fillport are no longer available, remove the existing
fillport by following the instructions below and install new style fillport.
1. Remove the locknut connecting the oxygen line to the connector fillport.
2. Remove the quick-release pin.
3. Remove the mounting screw connecting the flange to the base.
4. Remove the connector fillport from the base.
5. Remove the locknut and the two washers at the bottom of the connector fillport.
6. Remove the flange from the connector fillport assembly.
7. Remove the existing lanyard from the shaft.
8. Install new style fillport. Refer to Removing/Installing Connector Fillport Dust Cover
and Lanyard - Units Manufactured After 10/02/2005 on page 39.