SECTION 8—WIRING ASSEMBLIES
6. Secure the new transformer to the base with the two mounting screws (Detail "A").
Securely tighten
7. Connect the transformer assembly wire harness to the circuit board (Detail "B").
8. Install the control panel. Refer to Removing/Installing Control Panel on page 25.
9. Install the cabinet. Refer to Removing/Installing the Cabinet on page 21
Mounting Screw
Transformer
Assembly
Wire Harness

Replacing Bleed Resistor Assemblies

NOTE: For this procedure, refer to FIGURE 8.3 on page 45.
1. Remove the cabinet. Refer to Removing/Installing the Cabinet on page 21.
2. Disconnect bleed resistor assembly 1 from the power inlet (Detail "A").
3. Disconnect the bleed resistor assembly 1 from the grounding clip on the motor
assembly (Detail "B").
4. Disconnect bleed resistor assembly 2 from the control panel (Detail "C").
5. Disconnect the bleed resistor assembly 2 from the grounding clip on the motor
assembly (Detail "B").
6. Connect one end of the new bleed resistor assembly 1 to the power inlet.
7. Connect the other end of the new bleed resistor assembly 1 to the grounding clip on
the motor assembly.
8. Connect one end of the new bleed resistor assembly 2 to the other connection point on
the grounding clip on the motor assembly.
9. Connect the other end of the new bleed resistor assembly 2 to the control panel.
10. Install the cabinet. Refer to Removing/Installing the Cabinet on page 21.
HomeFill™II Compressor
DETAIL "A"
FIGURE 8.2 Replacing Transformer Assembly
DETAIL "B"
Transformer Assembly
Wire Harness
Mounting
Screw
Circuit Board
Connector
44
Part No 1141487