Removing/Installing The Cabinet Filter Grid; Removing; Installing - Invacare HomeFill II Service Manual

Hide thumbs Also See for HomeFill II:
Table of Contents

Advertisement

SECTION 5—CABINET ASSEMBLY

Removing/Installing the Cabinet Filter Grid

NOTE: For this procedure, refer to FIGURE 5.2.
Tools required for this procedure are: Flat head screwdriver and .09 drill bit

Removing

1. Remove the cabinet. Refer to Removing/Installing the Cabinet on page 21.
2. Perform one of the following:
• For units that have plastic rivets, remove the plastic rivets by pushing the rivets
from the inside out with a .09 drill bit, proceed to STEP 3.
• For units without plastic rivets, proceed to STEP 3.
3. Using the flat head screwdriver, pry the lip of the cabinet filter grid out of the cabinet
(Detail "C").

Installing

NOTE: For units that have plastic rivets, the cabinet MUST be replaced to use the new filter grid.
1. Install the cabinet filter grid (Detail "B") to the inside of the cabinet. Ensure that the
lip of the cabinet filter grid is on the outside of the cabinet.
2. Install the cabinet. Refer to Removing/Installing the Cabinet on page 21.
DETAIL "A"
Plastic Rivets
FIGURE 5.2 Removing/Installing the Cabinet Filter Grid
HomeFill™II Compressor
DETAIL "B"
Cabinet Filter
Grid
Lip
DETAIL "C"
NOTE: Lip of the filter grid MUST
be on the outside of the cabinet to
secure the filter grid in place.
22
Lip
Part No 1141487

Hide quick links: