SECTION 7—BASE COMPONENTS
CAUTION
Be careful not to cut the tubing when performing the following step.
2. Using the side cutters, cut the tie-wrap securing the tubing to the O
connector.
2
3. Remove the tubing from the O
connector.
2
4. Using a 5/8-inch wrench, remove the locknut securing the connector to the base.
5. Remove the connector and dust cap from the base.
6. Install the connector and dust cap to the base.
7. Secure the connector to the base with the locknut. Torque the locknut to 8-10 inch-
pounds.
8. Secure the tubing to the O
connector with a tie-wrap.
2
9. Install the cabinet. Refer to Removing/Installing the Cabinet on page 21.
Tubing
Locknut
(Not Shown)
Dust Cap
Base
O
Connector
2
Tie-Wrap
FIGURE 7.4 Replacing O
Inlet Connector/Inlet Connector Dust Cap
2

Replacing Transformer

NOTE: For this procedure, refer to FIGURE 7.5 on page 36.
Tools required for this procedure are:
• Phillips screwdriver
1. Remove the cabinet. Refer to Removing/Installing the Cabinet on page 21.
2. Note the four wire colors and sizes for the connectors located on the transformer.
3. Disconnect the four wire connectors from the transformer.
4. Remove the two mounting screws that secure the transformer to the base.
5. Remove the transformer from the base.
6. Install the new transformer into the base.
7. Secure the transformer to the base with two mounting screws. Securely tighten.
35