4-11. Rear Panel Assembly Access

Refer to Figure 4-4 during the following procedure:
1. Remove the six securing screws for the Rear Panel assembly housing.
2. Gently pull the rear panel housing from the Rear Panel.
3. Allow the rear panel housing to lay flat on the work surface by removing the two
4. Remove the two nuts at TB1 and TB2 on the paddle board; separate the associated
5. Remove P11 from J11. Then remove the two paddle board mounting screws and
6. Remove the jack screws for each connection on the rear panel housing, then gently
7. Reverse this procedure to install the Rear Panel assembly.

4-12. Front Panel Removal and Installation

Refer to Figure 4-5 during the following procedure:
4-6
1
REMOVE FACK SCREWS TO SEPARATE THE REAR PANEL CIRCUIT BOARD FROM
THE METAL HOUSING.
Figure 4-4. Rear Panel Assembly Access
ribbon cables from the Rear Panel board.
wires from the paddle board.
separate the paddle board from the Rear Panel assembly.
lift the Rear Panel assembly out from the housing.
REMOVE
1
1
1
1
REMOVE
F4-3.EPS